Physical dimensions:
Without external paddles 295mm
Environmental and physical requirements:
- Air temperature range: 20.0 - 39.0°C 3
- Skin temperature range:4.0 - 38.0°C
- Accuracy air/skin control: +/- 0.2°C
- Display range air/skin: 20.0 - 45.0°C
- 3.5°C key to adjust air/skin higher than 37.5°C
Oxigen administration
21% to 80% through one inlet
Humidification
Greater than 60% relative humidity with disposable sponge
Visibility
Hoods are plexiglass on all sides providing 100% visibility
Accessibility
- Access doors: two
- Mattress: Slide out mattress deck through head access door
- Access ports for I.V., respiration tubing and monitor leads: Yes
Display information
- Power inlet: Yes
- Battery charge: Yes
- Air control mode: Yes
- Skin control mode: Yes
- Key >37.5°C: Yes
- Four leds of indication of heating of the power 25%, 50%, 75%, 100%: Yes
Alarms
- High temperature of security limited: 40.0°C
- High air temperature: 1.5°C
- High skin temperature: 1.0°C
- Low air temperature: 3.0°C
- Low skin temperature: 1.0°C
- Failure air sensor: Yes
- Failure skin sensor: Yes
- Patient probe dislodgement: Yes
- Air flow: Yes
- Power failure: Yes
- Low battery: Yes
Electric specifications
- AC Input: 110/127V - 220V 50/60Hz 180w
- DC Input: 12VDC 115W
Battery
- Type: Sealed rechargeable
- Voltage: 12 VDC nominal
- Quantity: One
- Life expectancy: At least 200 cycles
- Autonomy: 4 hours
Physical specifications
- Width: 50 cm
- Depth: 170cm
- Height: 128cm
- Incubator’s weight: 71Kg
Security regulations
NBR-IEC 60601-1 NBR-IEC 60601-2-20 NBR-IEC 60601-1-2
For additional information please see brochure |